Compression and system movements
4.1
Testing the "compression" function
•
Attach the spring scale as shown in
compression plate during the following tests:
Tolerance: ± 1 kp.
4.1.1
Compression cutoff
•
System ON
PMF
Presetting
•
Select any value (e.g. 6 kp) on the potentiometer
the compression unit cut off at that setting?
SIE
Max. value for cutoff
•
Select the maximum value (20 kp) with the potentiometer
Does the unit cut off when it reaches the maximum value?
•
Correct display (chapter 2 "System",
PMF
OPCOMP
•
Position your forearm between the object table and the compression plate and check
the OPCOMP function.
Fig. 1:
MAMMOMAT Novation
Compression and system movements
Typically, the cutoff value is between 6 and 10 kp with the factory setting.
Compression
SPB7-250.831.01.02.02
08.04
(1/Fig. 1 / p. 18)
and make sure not to damage the
(2/Fig. 1 / p. 18)
(6/Fig. 1 / p.
13)).
Page 18 of 32
CS SD 24
and compress. Does
(2/Fig. 1 / p.
18). (20 kp).