Compression and system movements
4.1

Testing the "compression" function

Attach the spring scale as shown in
compression plate during the following tests:
Tolerance: ± 1 kp.
4.1.1

Compression cutoff

System ON
PMF
Presetting
Select any value (e.g. 6 kp) on the potentiometer
the compression unit cut off at that setting?
SIE
Max. value for cutoff
Select the maximum value (20 kp) with the potentiometer
Does the unit cut off when it reaches the maximum value?
Correct display (chapter 2 "System",
PMF
OPCOMP
Position your forearm between the object table and the compression plate and check
the OPCOMP function.
Fig. 1:
MAMMOMAT Novation
Compression and system movements
Typically, the cutoff value is between 6 and 10 kp with the factory setting.
Compression
SPB7-250.831.01.02.02
08.04
(1/Fig. 1 / p. 18)
and make sure not to damage the
(2/Fig. 1 / p. 18)
(6/Fig. 1 / p.
13)).
Page 18 of 32
CS SD 24
and compress. Does
(2/Fig. 1 / p.
18). (20 kp).