System
2-
2
System
2.1
Checks
PMP
Radiation protection
•
Check the radiation protection shield for possible damage
•
Information regarding the lead equivalent value must be legible.
PMP
Face shield and compression plate
•
Check the face shield and compression plate for correct seating and for damage
(2/Fig. 1 / p.
SIM
Level
•
Verify that the system is leveled with respect to the floor. Check the leveling screws
(3/Fig. 1 / p.
SIM
Swivel arm attachment
•
Check the swivel arm attachment by pulling on both support rails (minimum play of ap-
proximately 2° is correct).
SIM
Basic table
•
Check the mounting of the basic table as well as the locking mechanism for the expo-
sure system
Fig. 1:
Siemens AG
Medical Solutions
13).
13)with a water level and, if available, the floor mounting (optional).
(4/Fig. 1 / p.
(5/Fig. 1 / p.
13).
MAMMOMAT Checks
SPB7-250.831.01.02.02
08.04
(1/Fig. 1 / p.
13).
Page 13 of 32
CS SD 24
13
13).
MAMMOMAT Novation