Miscellaneous
6.2

Checks

SIE
Emergency STOP
Press Emergency STOP, see chapter "Compression and system movements"
(4/Fig. 1 / p.
- Are compression, rotation and vertical travel of lifting carriage blocked?
Release Emergency STOP.
PMF
Indicators
Check the LEDs on the operating console by starting the Panel Test via the service PC.
PMP
Error memory
Read out the error memory and the exposure counter with the service PC.
PMP
Record error memory
PMP
Delete the error memory
Enter the data of the exposure counter and the error memory in the protocol and delete
the error memory o n l y.
PMF
Auxiliary voltages
With the voltmeter, measure the auxiliary voltages according to circuit diagram
(SPB7-250.844...).
System OFF
PMF
OPDOSE
Check the selection of various anodes / filter combinations and OPDOSE according to
the Installation Start-Up instructions, chapter "Testing and adjusting the OPDOSE".
PMA
UIs and SPEED Infos
Check whether all relevant UIs and SPEED Infos have been completed.