Quality - Siemens MAMMOMAT Novation Maintenance Instructions Manual

Table of Contents

Advertisement

30
6.3

Quality

Proceed with the test according to the Quality Control Manual.
Q
Monitor check and viewing conditions
Perform the check according to the Quality Control Manual, chapter "Monitor check and
viewing conditions", to assess the quality of the acquisition workstation monitor.
Q
Printer check
Perform this check according to the Quality Control Manual, chapter "Printer check", to
assess the quality of the printer.
QIQ
Phantom image quality check
Perform the image quality check according to the Quality Control Manual, chapter
"Phantom image quality".
X
QIQ
Detector uniformity
Perform the image quality check according to the Quality Control Manual, chapter "De-
tector uniformity", to ensure that the pixel values of the detector do not deviate more
than 5% from each other.
X
QIQ
Beam quality (HVL)
Perform the image quality check according to the Quality Control Manual, chapter
"Beam quality", to optimize the dose calculation
X
QIQ
AEC image stability, repro and signal-to-noise (SNR)
Perform the check according to the Quality Control Manual, chapter "AEC image stabili-
ty, repro and signal-to-noise (SNR)", to assure, that the coefficient of variation for the
entrance dose and mAs is less than 0.05 and the SNR does not vary more than ± 15%
from the mean value.
X
QIQ
Ghost image evaluation
Perform the check according to the Quality Control Manual, chapter "Ghost image eval-
uation", to ensure that the ghosting is within an acceptable level.
X
MAMMOMAT Novation
SPB7-250.831.01.02.02
08.04
Miscellaneous
Page 30 of 32