22
4.4
Checking miscellaneous system movements
SIE
Rotation cutoff
•
Check whether the motorized rotation movements cut off when a preset angle is
reached. The angle (e.g. 90°) is preset with the potentiometer,
•
Check the automatic cutoff function when passing the 0° position as well.
PMF
Rotation movements
•
Is the rotation smooth and noiseless through the entire range?
•
Is the display correct (chapter 2 "System",
SIM
Blocking the rotation and vertical travel
•
Rotation and vertical travel must be blocked at a compression force of > 3 kp.
MAMMOMAT Novation
Compression and system movements
SPB7-250.831.01.02.02
08.04
(3/Fig. 1 / p.
(6/Fig. 1 / p.
13))?
Page 22 of 32
CS SD 24
18).
Siemens AG