" will flash, press the MEM button to switch between the alarm being OFF
5. Next "
or ON. To set the alarm, select "
to flash, press the MEM button to change the HOUR digits. Press the SET button to confirm.
6. Next the MINUTE digits will flash, follow the steps above to set the digits. Once Alarm 1 is
set you can also set Alarm 2 following the same steps. (See section 3.12)
7. After the minutes have been set, "CL" will appear on the display. Press the SET button to
disregard. See section 3.10 to clear the memory.
NOTE: In order to change any setting you must repeat the process and confirm each setting by pushing the
SET button.
3.3 Obtaining Accurate Measurements
Your blood pressure can vary based on numerous factors, physiological conditions, and your surroundings.
Follow these guidelines to obtain accurate and error-free measurements of your blood pressure and pulse rate.
3.3A Before Measuring
In the morning before breakfast,
2 hours after dinner, before taking
medication.
Avoid coffee and smoking within the
hour, and no exercise 30 minutes
before measuring.
Do not speak while taking the
measurement.
Sit with legs uncrossed so as not to
restrict blood flow.
Ensure that the BP monitor is level
with the heart while the arm is
supported on the table.
3.3B Common Sources of Error
All efforts by the patient to support the arm can increase the blood pressure. Make sure you are in a comfortable,
relaxed position and do not activate any of the muscles in the measurement arm during the measurement. Use a
cushion for support if necessary.
ATTENTION!
Comparable and reliable blood pressure measurements always require the same conditions with a
peaceful and calm environment. Ensure that you take measurements under the same conditions to obtain
an accurate estimate of blood pressure variation patterns.
• If the arm artery lies considerably lower or higher than the heart, an erroneous value of blood pressure is
measured. Each 15 cm difference in height results in a measurement error of 10 mmHg.
• A loose cuff causes false measurement values.
• With repeated measurements, blood accumulates in the arm, which can lead to false results. Consecutive blood
pressure measurements should be repeated after at least a 15 second pause or after the arm has been held up in
order to allow the accumulated blood to flow away.
" and press the SET button. The HOURS will begin
Empty bladder (if necessary).
Rest quietly for 5 minutes. Remain
calm and quiet while the measurement
is in process.
Take measurements on the non
dominant arm.
Sit with back supported and
measurement arm resting on a table.
Sit with feet flat on the floor.
Need help?
Do you have a question about the PROTOCOL BD352 and is the answer not in the manual?