Connecting An Oximax Sensor - Nellcor OxiMax N-600X Operator's Manual

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

Setting Up the Monitor
Connecting an O
26
M
Sensor
XI
AX
The O
M
sensor type is shown at the bottom of the display when
XI
AX
an O
M
sensor is connected to the N-600x or when the monitor
XI
AX
completes POST with an O
Note: Sensor LED light emissions fall within Class 1 level, according
to IEC 60825-1:2001.
Caution: Use only Nellcor-approved O
oximetry cables.
Note:Physiological conditions, medical procedures, or external
agents that may interfere with the monitor's ability to detect and
display measurements include dysfunctional hemoglobin, arterial
dyes, low perfusion, dark pigment, and externally applied coloring
agents, such as nail polish, dye, or pigmented cream.
1. Firmly connect a DOC-10 pulse oximetry cable to the SpO
O
M
Sensor Port of the monitor.
XI
AX
2. Connect a Nellcor O
DOC-10 pulse oximetry cable.
M
sensor attached.
XI
AX
M
sensors and pulse
XI
AX
M
SpO
sensor to the opposite end of the
XI
AX
2
N-600x Operator's Manual
2