Principles Of Operation; Overview - Nellcor OxiMax N-600X Operator's Manual

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

P r i n c i p l e s o f O p e r a t i o n

Overview

N-600x Operator's Manual
The N-600x uses pulse oximetry to measure functional oxygen
saturation in the blood. Pulse oximetry works by applying an O
sensor to a pulsating arteriolar vascular bed, such as a finger or toe.
The O
M
sensor contains a dual light source and a photo detector.
XI
AX
Bone, tissue, pigmentation, and venous vessels normally absorb a
constant amount of light over time. The arteriolar bed normally
pulsates and absorbs variable amounts of light during the pulsations.
The ratio of light absorbed is translated into a measurement of
functional oxygen saturation (SpO
Because a measurement of SpO
O
M
sensor, excessive ambient light can interfere with this
XI
AX
measurement.
Specific information about ambient conditions, O
application, and patient conditions is contained throughout this
manual.
Pulse oximetry is based on two principles: that oxyhemoglobin and
deoxyhemoglobin differ in their absorption of red and infrared light
(spectrophotometry), and that the volume of arterial blood in tissue
(and hence, light absorption by that blood) changes during the pulse
(plethysmography). A pulse oximeter determines SpO
red and infrared light into an arteriolar bed and measuring changes in
light absorption during the pulsatile cycle. Red and infrared
low-voltage light-emitting diodes (LED) in the oximetry O
sensor serve as light sources; a photo diode serves as the photo
detector.

Principles of Operation

).
2
is dependent upon light from the
2
M
XI
AX
by passing
2
M
XI
AX
sensor
M
XI
AX
171

Advertisement

Table of Contents
loading

Table of Contents