NELLCOR SENSORS
Selecting a Sensor
Cleaning and Reuse
Performance Considerations
SELECTING A SENSOR
WARNING: Use only NELLCOR oxygen transducers. Use
of other oxygen transducers may cause improper
oximeter performance.
WARNING: Before use, carefully read sensor directions
for use.
NELLCOR sensors are briefly described in Table 1. To select
the appropriate sensor, consider the patient's weight, activity
level, perfiision, which sensor sites are available, whether
sterility is required, and the duration of monitoring.
Table 1: NELLCOR Sensors
Sensor
OXISENSOR oxygen transducers
(Sterile, single-use only)
OXIBAND oxygen transducers
(Reusable with disposable nonsterile
adhesive)
DURASENSOR oxygen transducer
(Reusable, nonsterile)
NELLCOR reflectance oxygen
transducer (Nonsterile)
Patient
Model
Size
N-25
<3 or >40 kg
1-20
1-20 kg
D-20
10-50 kg
D-25(L)
>30kg
R-15
>50kg
OXI-A/N
<3 or >40 kg
OXI-P/I
3-40 kg
DS-100A
>40kg
RS-10
>40kg
21