Remark :
During execute the Manual Datalogger, press the " ▲
Button " ( 3-4, Fig, 1 ) or "
set the measuring Location no. ( 1 to 99, for example
room 1 to room 99 ) to identify the measurement location
, the lower Display will show P x ( x = 1 to 99 ).
b. Finish the Datalogger
Press the " LOGGER Button ( 3-9, Fig. 1 ) > 2 seconds
continuously, the " DATA RECORD " indicator will be
disappeared and finish the Datalogger.
5-4 Peak hold record mode
( Fast Datalogger )
1) Recommend the Force gauge install to the Test stand
( such as FS-1001, FS-1002.. ), Power on the meter.
2) Press the " PEAK Button " ( 3-6, Fig. 1 ) once, the Display
will the indicator " PEAK " and " FAST ".
3) Press the " Zero Button " ( 3-4, Fig. 1 ) once, the Display
will set to " ZERO " value.
4) * Press the " LOGGER Button ( 3-9, Fig. 1 ) once , the
LCD will show the indicator " DATA RECORD " at the
same time the sampling time value in ms ( milli
seconds ) will appear a while in the bottom LCD, for
example " 10 ". Now the meter is ready for the " Peak
hold record mode ( Fast Datalogger measurement ) ".
* If apply the force value ( tension/compression ) > 0.2
Kg ( 1% full capacity ), then the measurement value
according the sampling period will save to the internal
memory circuit in sequence.
Remark :
How to set the sampling time, refer to Chapter 7-4
*
page 22.
Button " ( 3-7, Fig. 1 ) to
14