Post-Installation Check - Endress+Hauser Magphant Operating Instructions Manual

Electromagnetic flowmeter
Hide thumbs Also See for Magphant:
Table of Contents

Advertisement

Installation
20
 8
Mounting the adapter piece on the sensor
1
Adapter piece made of stainless steel, 1.4435 (F316L)
2
Plastic union nut
3
Metal union nut
6.3

Post-installation check

Is the device undamaged (visual inspection)?
Does the measuring device conform to the measuring point specifications?
For example:
• Process temperature
• Process pressure (refer to the section on "Pressure-temperature ratings" in the "Technical Information"
document)
• Ambient temperature
• Measuring range
Is there sufficient distance between the sensor and the next pipe bend?
Is the electrode axis at a 90° angle to the flow direction?
Is the sensor area fully immersed in the liquid?
Has the correct orientation for the sensor been selected?
• According to medium temperature
• According to medium properties (risk of air pockets and build-up of solids)
Is the measuring device adequately protected against precipitation and direct sunlight?