Charder MS2510 Calibration Procedure

Charder MS2510 Calibration Procedure

Advertisement

CALIBRATION PROCEDURE
To calibrate the scale, please turn off the scale.
Operation
Display
Turn the scale on using the ON/OFF key.
Press UNIT key for 3 seconds and SETUP
will show shortly. UNIT will appear on the
display.
Press the TARE key repeatedly until the
"CAL ib" symbol is displayed.
Press the HOLD key.
Press the TARE key.
The triangle [
] will display on the top of
left hand side of the display. If not, press
the TARE key.
Press the HOLD key repeatedly until the
"CAL 0" symbol is displayed.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MS2510 and is the answer not in the manual?

Questions and answers

Summary of Contents for Charder MS2510

  • Page 1 CALIBRATION PROCEDURE To calibrate the scale, please turn off the scale. Operation Display Turn the scale on using the ON/OFF key. Press UNIT key for 3 seconds and SETUP will show shortly. UNIT will appear on the display. Press the TARE key repeatedly until the “CAL ib”...
  • Page 2 Press the TARE key, the display will show present ZERO Count value (6000~12000). Then press the UNIT key. Press the HOLD key. Press the TARE key. Enter the calibration weight (200 kg). Press TARE key and HOLD key to set the weight. Confirm by pressing the UNIT key and the number on the display will be 000.0...