CLEANING THE BRACKET SYSTEM

CLEANING PROCESS

WARNING
Follow your EMS service's protocol or medical device manufacturer's User Guide for the recommended cleaning
procedure of your medical devices. Cleaning procedure below does NOT provide recommendations for cleaning of the
medical device. The following cleaning guidelines consider the Technimount mounting assemblies and components.
Remove medical device from the bracket if thorough cleaning is needed.
The Bracket can be cleaned by using a pressure washer. The bracket may show some signs of oxidation or discoloration
from continuous washing, however, degradation of the cot/stretcher's performance or functionality will not be
affected due to power washing if cleaning instructions are properly followed.
The material used to manufacture the Bracket is high-grade aluminum, it is very durable and resistant to corrosion,
however high levels of sodium or a high concentration of sodium or acid may prematurely corrode the product.
Always rinse with soft soap and water if exposed.
9 Use of a power washer can accelerate removal of contaminants collected during the use of the product.
9 Rinse with clean water.
9 Towel dry all components.
9 Allow all other components to air dry.
9 Avoid over saturation and ensure that the product does not stay wet longer than the cleaner manufacturer's
guidelines for proper disinfecting.
WARNING
9 When cleaning, always use appropriate personal protection equipment (PPE) based on established protocols (e.g.,
gloves, eye wear, etc.).
CAUTION
9 Do not steam clean or use ultrasonic cleaners on the system or any of its components.
9 Do not immerse the metal parts/components in water.
9 Maximum water temperature should not exceed 180°F/82°C.
9 Maximum water pressure should not exceed 1500 psi/103.5 BAR . If a pressure washer is being used to clean the
unit, the pressure nozzle must be kept a minimum of 24 inches (61 cm) from the unit.

CLEANING SOLUTIONS

Phenolic type or quaternary type disinfectants can be used (excluding Virex® TB). Iodophor type disinfectants,
Peracetic Acid, Chlorine, and especially Peroxide Oxygen are not recommended for use because staining and corrosion
may result.
Suggested cleaners for the Bracket:
9 Quaternary Cleaners (active ingredient - ammonium chloride)
9 Phenolic Cleaners (active ingredient - o-phenyl phenol)
9 Chlorinated Bleach Solution (3.25% - less than 1-part bleach to 100 parts water)
Return To Table of Contents
26

Maintenance Guide

BP25 and 25-HD UG 202111-01