Download Print this page

Thermo Scientific 42iQLS Instruction Manual page 77

Low source no-no2-nox analyzer

Advertisement

Reaction Chamber
Thermo Scientific
The Reaction Chamber screen provides status and alarms related to the
reaction chamber module. If an item being monitored goes outside the
lower or higher alarm limit, an alarm is activated.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Reaction
Chamber
The Reaction Chamber screen contains the following information:
Across:
Chamber: This column lists items associated with the reaction
chamber.
Value: Displays the current value for each item.
Low Alarm: Displays low alarm status for each item.
High Alarm: Displays high alarm status for each item.
Units: Displays units for each item.
Down:
Instrument Temperature: Displays the current instrument
temperature reading.
Flow: Displays the current sample flow reading.
Chamber Pressure: Displays the current chamber pressure reading.
Chamber Temperature: Displays the current chamber temperature
reading. User can adjust low and high alarm limits.
Chamber Temp Sensor Short: Displays OK/Fail for chamber
temperature sensor short.
Operation
Settings
42iQLS Instruction Manual
3-51

Advertisement

loading