Focusing Arm: Configuration 1 for Table Inspections

 
In this configuration, the focusing arm
is inserted from below. If necessary, the
focusing arm can be set in a slanted position,
for example if the specimen cannot be positi-
oned horizontally under the microscope.
 
For the greatest possible stability, you
should extend the small horizontal arm
as little as possible in order to reduce a lever
effect.
Assembly
1. Push the holder of the focusing arm from
below into the opening of the small hori-
zontal arm.
Leica F12 I Floor Stand
2. Tighten the locking screw.
 
For this configuration, the counter-
screw must be locked tight with the
washer at all times. Otherwise the entire micro-
scope could fall onto the specimen!
3. Be sure to secure the focusing arm using
the screw provided!
Manual
4. Unscrew both clamping screws on the focu-
sing arm. Modify the tilt of the focusing arm
and tighten the locking screw.
17