User Manual
10 744 763 - Version 06

Advertisement

Table of Contents
loading

Summary of Contents for Leica FL800 ULT

  • Page 1 Leica FL800 ULT User Manual 10 744 763 - Version 06...
  • Page 2 The information provided by this manual is directly related to the operation of the equipment. Medical decision remains the responsibility of the clinician. Leica Microsystems has made every effort to provide a complete and clear user manual highlighting the key areas of product use. Should additional information regarding the use of the product be required, please contact your local Leica representative.
  • Page 3: Table Of Contents

    Care and maintenance What to do if...? General Leica FL800 ULT Technical data Technical data Leica FL800 ULT Compatibility Ambient conditions Standards fulfilled Manufacturer's declaration of electromagnetic compatibility (EMC) Leica FL800 ULT / Ref. 10 744 763 / Version 06...
  • Page 4: Introduction

    Please contact your local The Leica FL800 ULT is an accessory for Leica surgical microscopes. representative for availability. In this user manual the functions of the Leica FL800 ULT are described. The subsequent screenshots are reference only and not For information and description of the Leica surgical microscope, representative for all models.
  • Page 5: Safety Notes

    Intended use Signs and labels • The Leica FL800 ULT is a surgical microscope accessory used in Type label viewing intra-operative blood flow in the cerebral vascular area, blood flow of the coronary vascular and bypass grafts during coronary artery bypass (CABG) surgery, as well as blood flow during plastic and reconstructive surgery.
  • Page 6: Description

    The following is an example of a Leica surgical microscope. Function The illumination of the Leica FL800 ULT is a xenon lamp and is located in the Leica surgical microscope. This lamp provides visible light and the Leica FL800 spectrum. This image cannot be observed through the surgical microscope, but it is recorded using a special camera and visualized on the monitor.
  • Page 7: Controls

    LED lights up red = recording in progress X– Step Loop Y– 3.3.3 Recording system unit (optional) Please refer to the user manual of the corresponding recording system unit. Leica FL800 ULT / Ref. 10 744 763 / Version 06...
  • Page 8: Preparation Before Surgery

    You see an overview with which buttons of the left handle the Leica FL800 functions can be operated. Select the "Fluorescence Vascular FL800" user preset. Click on the "Select" button. Leica FL800 ULT / Ref. 10 744 763 / Version 06...
  • Page 9 Open the "Handle Left" or "Handle Right" tab to assign FL800 functions to the handle. Brightness Open the "FL" tab to set the brightness in FL800 mode to an adequate level. Leica FL800 ULT / Ref. 10 744 763 / Version 06...
  • Page 10 Preparation before surgery Video Fine Focus The Leica FL800 ULT offers fine focusing and parfocality reset of the video focus. The video focus can be adapted to your needs by pressing the focus button up (3) or/and down (1). This command can be given to the GUI and from the handle, if defined.
  • Page 11: Checking The Leica Fl800 Ult Illumination

    • With low intensity, use the back side Leica FL800 ICG Fluorescence Standard Please read the Leica FL800 user manual! high intensity 10 713 880-xx 1 ICG test pattern 1 ICG test pattern Leica FL800 ULT / Ref. 10 744 763 / Version 06...
  • Page 12: Pre-Operation Checklist (Leica Fl800 Ult)

    Check optical accessories for cleanliness. Remove dust and dirt. Leica FL800 ULT application When using Leica FL800 ULT please ensure to have a Doppler Ultrasound or similar in place, in case of none or insufficient blood flow visualization out of the ICG/Leica FL800 ULT procedure is given.
  • Page 13: Operation

    Press the joystick to the left to switch between the modes. WARNING Danger of injury to the patient due to excessive Leica FL800 radiation Avoid extended and/or excessively frequent use of Leica FL800 radiation. Leica FL800 ULT / Ref. 10 744 763 / Version 06...
  • Page 14: Care And Maintenance

    • Are the power cables attached correctly? • Are all connecting cables attached correctly? • Are all video cables attached correctly? General For malfunctions regarding the Leica surgical microscope refer to the user manual of the Leica surgical microscope. Leica FL800 ULT Malfunction Cause...
  • Page 15: Technical Data

    The figure shows the camera image size with respect to the field of view for the visual video camera and the Leica FL800 ULT NIR camera. Please be aware that the field of view is not fully covered by the documentation system.
  • Page 16: Manufacturer's Declaration Of Electromagnetic Compatibility (Emc)

    (EMC) The Leica FL800 ULT was tested in combination with Leica surgical microscopes. For the EMC declaration, please refer to the user manual of the Leica surgical microscope. Leica FL800 ULT / Ref. 10 744 763 / Version 06...
  • Page 17 Leica FL800 ULT / Ref. 10 744 763 / Version 06...
  • Page 18 10 744 763en/06• Copyright © by Leica Microsystems (Schweiz) AG, Medical Division, CH-9435 Heerbrugg, 2019 • Printed –10.2019– Subject to modifications. • LEICA and the Leica Logo are registered trademarks of Leica Microsystems IR GmbH.