Security Seal Installation - Detecto MedVue MV1 Owner's Manual

Medical weight analyzer
Table of Contents

Advertisement

SECURITY SEAL INSTALLATION

If your metrology laws require the use of physical sealing, a lead and wire security seal can be
installed on the MV1 to prevent the end cap from being removed from it to gain access to the
calibration switch, thereby preventing unauthorized access to the calibration adjustments.
NOTE: In order to install the lead and wire security seal, the two screws securing the left
end cap (as viewed from the MV1 front) MUST be replaced with two drilled screws. They
are available from the Cardinal/Detecto Parts Department. The part number is MVSK
(MEDVUE SEALING KIT).
To install a lead and wire security seal, refer to
Figure No. 12 and follow the steps provided below:
1. Remove the two screws securing the left end
cap (as viewed from the MV1 front).
2. Install the two drilled screws (and lock
washers) in the end cap making certain they
are securely tightened before proceeding.
3. Thread the sealing wire through the hole in
the drilled screws.
4. Pull the wire tight and install the lead seal.
5. The screws cannot be removed without
damaging the seal.
8555-M512-O1 Rev E MedVue MV1 Owner's
Figure No. 12
49