With SCALE displayed, press the ENTER key. The display will change to CAPACITY. Proceed
to the CAPACITY (Scale Capacity) parameter.
Otherwise, press the ID/HEIGHT key to advance to the
CAPACITY (Scale Capacity)
The display will show CAPACITY and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 1 through 65,535.
OIML (OIML European Requirements)
The display will show OIML and the current setting. If the value displayed is acceptable,
press the ENTER key to save it. Otherwise, use the numeric keys 0/NO or 1/YES to select a
new setting, and then press the ENTER key to save it.
Allowable values are YES or no.
YES = OIML (European) specific
requirements are enabled
INTERVAL (Interval Setting)
The display will show INTERVAL and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 1, 2, or 5.
dECIMAL (Decimal Point Position)
The display will show dECIMAL and the current setting. If the value displayed is acceptable,
press the ENTER key to save it. Otherwise use the numeric keys to enter a new value, and
then press the ENTER key to save it.
Allowable values are 0, 1, 2, or 3.
0 = # # # # # #
MOTION RANGE (Motion Range)
The motion range is the number of divisions of change permitted before indicating unstable
(the STABLE annunciator turns off).
The display will show MOTION RANGE and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 0 through 9.
SAMPLE RATE (Sample Rate)
The display will show SAMPLE RATE and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
The sample rate may be set from a minimum of 1 sample per second to a maximum of 100
samples per second in one sample per second intervals.
Allowable values are 1 through 100.
CALIBRATION SETUP, CONT.
1 = # # # # #. #
8555-M512-O1 Rev E MedVue MV1 Owner's
ZERO
menu.
no = OIML (European) specific
requirements are disable
2 = # # # #. # #
3 = # # #. # # #
25