Single Calibration Point
L L OAdEd CAL WT (Load Calibration Weight)
The display will show LOAdEd CAL WT (which is a prompt for entry of the calibration weight
value) and placement of this amount of test weight on scale platform.
1. Make certain scale platform is empty and free of debris.
2. When the display is showing LOAdEd CAL WT, the current setting for the calibration
weight will be displayed also.
3. If the value displayed is acceptable, place the calibrated test weight on the scale
platform, and then press the ENTER key. Proceed to Step 6.
4. Otherwise, determine the exact amount of test weight to be placed on the scale
platform, and then using the numeric keys enter this value. Allowable values are 1
through scale capacity (refer to the capacity label on MV1).
NOTE: A minimum of 50% of the scale's capacity is required. However, 70% to 100% is
recommended.
5. Verify that the numbers entered are the same as the amount of test weight, and then
press the ENTER key.
6. The display will change to show PLEASE WAIT and then starting at the left and
proceeding right, a series of dashes will appear on the display, and then starting at the
right and proceeding left, disappear. Next, the display will change to UNLOAd SCALE.
UNLOAd SCALE (Unload Scale)
With the display showing UNLOAd SCALE:
1. Remove the weight from the scale platform.
2. Press the ENTER key.
3. The display will change to show PLEASE WAIT and starting at the left and proceeding
right, a series of dashes will appear on the display, and then starting at the right and
proceeding left, disappear.
4. Next, the display will change to show the GRAVITY ENAbLE prompt.
CALIBRATION SETUP, CONT.
8555-M512-O1 Rev E MedVue MV1 Owner's
29