Calibration Setup (Protected Setup Parameters) - Detecto MedVue MV1 Owner's Manual

Medical weight analyzer
Table of Contents

Advertisement

Your MedVue Model MV1 Medical Weight Analyzer has been thoroughly tested and calibrated
before being shipped to you. If you received it with a scale, calibration is not necessary. If the
MV1 is being connected to a scale for the first time or recalibration is necessary for other
reasons, the following describes the steps to perform calibration setup on the MV1.
To Enter Calibration Setup:
1. With the power off, remove the two (2) screws from the left
end cap (as viewed from the front of the MV1) and remove
the end cap.
2. Locate the calibration switch (access is through a hole near
in the side panel). See Figure No. 11.
3. Press the ON/OFF key to turn the MV1 on.
4. The MV1 will show the software version and then change to
the Gross weight mode.
5. With the MV1 on, press and hold the calibration switch until
the display changes to show:
SEtuP
SCALE
6. Release the calibration switch.
7. The MV1 is now ready for setup and calibration
During setup and calibration, it is necessary to enter operational parameters via the MV1's
keypad. The following keys and the numeric keypad are used to input the values.
NOTE: The membrane keypad is not to be operated with pointed objects
(pencils, pens, fingernails, etc.). Damage to keypad resulting from this
practice will NOT be covered under warranty.
Navigation Keys
The Navigation Keys are used to move around the menu selections during setup and
calibration. They are used to select the parameter to change, and to advance to the next
parameter, backup to the previous parameter, or to "exit" to the previous screen.
GROSS/NET – (Previous Menu or Parameter)
If at the menu level, this will take you to the next menu item in the list below. If
TIME SETTINGS is currently displayed and the setup has been accessed
through the calibration switch, then this will return you to SCALE. If setup has
been accessed through setup review, then pressing this when
TIME SETTINGS is displayed will return you to SYSTEM. If at the parameter
level, the exact item this takes you to may depend on a [parameter condition].
CALIBRATION SETUP
(Protected Setup Parameters)
CAL
8555-M512-O1 Rev E MedVue MV1 Owner's
Calibration
Switch
access hole
Figure No. 11