SERIAL AND USB DEVICE PORT COMMANDS, CONT.
<LF>P<CR>
Weight analyzer will respond with all information formatted with each item on its own line.
<LF>J<CR>
Weight analyzer will respond with all information formatted with each item separated by
commas (comma delimited) followed by a carriage return.
<LF>Z<CR>
Zero the scale within the constraints of the zero settings.
<LF>T<CR>
Scale attempts to tare itself setting the tare weight to the current gross weight.
<LF>T<xxxxxx.xxx><CR>
Scale attempts to set the weight (represented as<xxxxxx.xxx> in the string) to the
requested value.
<LF>M<CR>
Scale returns the current tare weight.
<LF>U<CR>
Scale will toggle between calibration units if conversion units have been set up in scale
calibration.
<LF>A<CR>
Scale will respond with:
<LF>MedVue<CR>
<LF>XH<CR>
Scale will respond with the current height:
<LF>Height<CR>
<LF>XB<CR>
Scale will respond with the current BMI:
<LF>BMI<CR>
<LF>XI<CR>
Scale will respond with the currently entered ID:
<LF>ID<CR>
<LF>XA<CR>
Scale will respond with current weight, height, BMI, and ID in comma delimited format:
<LF>
weight (see <LF>W<CR> above), Height, BMI, ID<CR>
NOTE: If the MedVue weight analyzer has been set to continuous mode of outputting weight, it
will output the weight in the format listed above for <LF>W<CR> every 500 milliseconds.