Tare Storage And Recall - Detecto MedVue MV1 Owner's Manual

Medical weight analyzer
Table of Contents

Advertisement

Tare Storage and Recall

NOTE: ID must be enabled to use stored Tare. Refer to SYSTEM Menu, Enable ID.
To Store Tare Weight
1. In Gross Weight mode (GROSS annunciator on), press the TARE key.
2. The Display will change to show ENTER TARE WT and the TARE annunciator will turn on.
3. Use the numeric keys and enter a tare weight. Note that when entering tare weights,
a maximum of 4 digits can be entered and the numbers advance from right to left in
the display. The number of leading or trailing zeros required to obtain your desired
tare is dependent upon the DIVISION VALUE selected in the setup procedure.
For example:
4. After the tare weight has been entered, press the ID/HEIGHT key.
5. Display will change to show
6. Using numeric keys, enter the memory location (1 to 99).
7. Press the ENTER key.
8. The tare weight is now stored in memory and the display will show a minus Net weight
and NET annunciator will turn on.
9. Assist patient onto scale.
10. When weight is stable (STABLE
11. If a printer is connected to scale, press the PRINT key to print a ticket.
12. Assist patient off scale.
To Recall Tare Weight
1. Press the ID/HEIGHT key.
2. Display will change to show
3. Using numeric keys, enter the memory location (1 to 99).
4. Press the TARE key.
5. The display will change to show ENTER TARE WT and the previously entered tare weight.
Note that the TARE annunciator will turn on.
6. If tare weight displayed is acceptable, press the ENTER key to save it.
7. Otherwise, press the ZERO key to clear the tare weight and return the MV1 to the Gross
weight mode.
OPERATION, CONT.
Division Value:
Desired Tare Value:
Key Sequence:
.
annunciator will turn on), read weight display.
.
8555-M512-O1 Rev E MedVue MV1 Owner's
0.2 lb
10.6 lb
1 0 6