Welch Allyn CSM 7100 Service And Repair Training page 61

Connex spot monitor

Advertisement

Initialize NIBP
• In WAST, double click the device and go to
verify and calibrate
• Select NIBP sensor and click Begin
• Click Initialize NIBP board and confirm
• Follow instructions on screen to perform:
– the Power Calibration Test – Primary/Safety
– the Pressure Calibration – Primary/Safety