Maintenance; Cleaning And Storage; Specifications - Extech Instruments CO210 User Manual

Co2 monitor and datalogger
Hide thumbs Also See for CO210:
Table of Contents

Advertisement

Maintenance

CLEANING AND STORAGE

Specifications

Find Quality Products Online at:
1. The meter should be cleaned with a damp cloth and mild detergent when necessary. Do not use
solvents or abrasives.
2. Store the meter in an area with moderate temperature and humidity.
Function
Range
CO
0 to 9999ppm
2
-10 to 60°C
Temperature
14 to 140°F
Humidity
0.1 to 99.9%
Display
LCD
Sensor Type
CO
: NDIR (non-dispersive infrared) technology
2
Humidity: Capacitance sensor;
Temperature (air): Thermistor
Response
CO
: <2min for 90% step change
2
Temp: <2min for 90% step change
%RH: <10min for 90% step change
Datalogging
Up to 15999 points
Sample rate
1 second to 4 hours-59 minutes-59 seconds
Operating Conditions
-10 to 60°C (14 to 140°F); < 90% RH non-condensing
Storage Conditions
-20 to 60°C (-4 to 140°F); <99% RH non-condensing
Power Supply
5VDC (±10%),
Dimensions / Weight
117x102x102mm (4.6x4x4"); 204g (7.2 oz.)
GlobalTestSupply
www.
Resolution
Accuracy
1ppm
±(5%rdg +50ppm)
0.1°
±0.6°C/0.9°F
±3%(10 to 90%)
0.1%
±5%(< 10% or > 90%)
500mA (AC adaptor supplied)
8
CO210-en-EU_V1.2 8/14
.com
sales@GlobalTestSupply.com