Spectrum Analyzer Hm5014-2; Specifi Cations - Hameg HM5014-2 Service Manual

Spectrum analyzer
Hide thumbs Also See for HM5014-2:
Table of Contents

Advertisement

1 GHz Spectrum Analyzer HM5014-2
Valid at 23 °C after a 30 minute warm-up period
Frequency Characteristics
Frequency Range :
0.15 MHz to 1.050 GHz
Stability:
± 5 ppm
Ageing:
± 1 ppm/year
Frequency Resolution:
1 kHz (6
Center Frequency Range:
0 to 1.050 GHz
LO Frequency Generation:
TCXO with DDS (Digital Frequency Synthesis)
Span Setting Range:
Zero Span and 1 MHz –1000 MHz
(1-2-5 Sequence)
Marker:
Frequency Resolution:
1 kHz, 6
Amplitude Resolution:
0.4 dB, 3
Resolution Bandwidths
(RBW) @ 6dB:
1 MHz, 120 kHz and 9 kHz
Video Bandwidth (VBW):
4 kHz
Sweep Time
(automatic selection):
40 ms, 320 ms,1 s*)
Amplitude Characteristics (Marker Related) 150 kHz – 1 GHz
Measurement Range:
-100 dBm to +10 dBm
Scaling:
10 dB / div., 5 dB/div.
Display Range:
80 dB (10 dB/div.),
40 dB (5 dB/div.)
Amplitude Frequency Response (at 10 dB Attn., Zero Span and RBW 1 MHz,
Signal – 20 dBm):
± 3 dB
Display (CRT):
8 x 10 division
Amplitude Scale:
logarithmic
Display units:
dBm
Input Attenuator Range:
0 – 40 dB (10 dB-increments)
Tolerance of input attenuator: ± 2 dB relative to 10 dB position
Max. Input Level (continuous)
40 dB attenuation:
+20 dBm (0,1 W)
0 dB attenuation:
+10 dBm
Max. DC Voltage:
± 25 V
Max. Reference Level: +10 dBm
Reference Level Accuracy rel. to 500 MHz, 10 dB Attn., Zero Span and
RBW 1 MHz:
± 1 dB
Min. Average Noise Level:
ca. -100 dBm (RBW 9 kHz)
Intermodulation Ratio
rd
(3
Order):
typical ›75 dBc (2 Signals: 200 MHz,
203 MHz, -3 dB below Reference Level)
Harmonic Distortion Ratio
(2
nd
harm.):
typical › 75dBc (200MHz, Reference Level)
Bandwidth Dependent Amplitude Error rel. to RBW
1 MHz and Zero Span:
± 1 dB
Digitization Error:
±1 digit (0.4 dB) at 10 dB/div. scaling
(Average, Zero Span)
Inputs/Outputs
Measuring Input:
Input Impedance:
50 Ω
VSWR: (Attn. ≥ 10 dB)
typ. 1.5:1
Tracking Generator Output: N-socket
Output Impedance:
50 Ω
Test Signal Output:
BNC-Buchse
Frequency, Level:
48 MHz, -30 dBm (± 2 dB)
Supply Voltage for Probes (HZ 530):
Audio Output (phone):
3.5mm Ø jack
RS-232 Interface:
9pol./Sub-D
Functions
Keyboard Input:
Center Frequency, Reference Level,
Tracking Generator Level
Rotary Encoder Input:
Center Frequency, Reference Level, Marker,
Tracking Generator Level
Max. Hold Detection:
Peak Value Acquisition
Quasi-Peak Detection:*
Quasi-Peak Valuation
Average:
Mean Value Acquisition
Ref. Spectrum Memory:
2 k x 8 bit
SAVE / RECALL:
Save and Recall of 10 Instrument Settings
AM demodulation:
for audio
LOCAL:
RS-232 Remote Control OFF
Readout:
Display of various Measurement Parameters
1
digit in readout)
2
1
digit,
2
1
digit
2
N socket
6 V DC
Tracking Generator
Frequency Range:
0.15 MHz to 1.050 GHz
Output Level:
-50 dBm to +1 dBm
Frequency Response (0.15 MHz – 1 GHz)
+1 dBm to -10 dBm:
± 3 dB
-10,2 dBm to -50 dBm:
± 4 dB
Digitization Error:
± 1 digit (0.4 dB)
Spurious Outputs
better than 20 dBc
General information
CRT:
D14-363GY, 8 x 10 cm with internal graticule
Acceleration Voltage:
approx. 2 kV
Trace Rotation:
adjustable on front panel
Ambient Temperature:
10° C to 40° C
Power Supply:
105-253 V, 50 / 60 Hz ± 10 %, CAT II
Power Consumption:
approx. 35 W at 230V/50 Hz
Safety Class
Safety Class I (EN61010-1)
Dimensions (W x H x D):
285 x 125 x 380 mm
Weight:
approx. 6.5 kg
*) in combination with software AS100E only
Accessories supplied: Line Cord, Operators Manual, HZ21 Adapter Plug
(N-plug with BNC socket) and Software for Windows on CD-ROM
Optional accessories:
HZ70 Opto-Interface (with optical fiber cable)
HZ520 Antenna
HZ530 Near Field Probe Set for EMI Diagnosis
S p e c i f i c a t i o n s
5
Subject to change without notice

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents