Spectrum Analyzer Hm5510; Specifi Cations - Hameg hm5510 User Manual

Spectrum analyzer
Hide thumbs Also See for hm5510:
Table of Contents

Advertisement

Available languages

Available languages

1 GHz Spectrum Analyzer HM5510
Specifi cations
(Valid at 23 degr. C after a 30 min. warm-up period.)
Frequency Characteristics
Frequency Range:
0.15 MHz to 1.050 GHz
Stability:
±5ppm
Ageing:
±1 ppm/year
Frequency Resolution:
1 kHz (6½ -digit in readout)
Center Frequency Range:
0 to 1.050 GHz
LO Frequency Generation:
TCXO with DDS (Digital Frequency Synthesis)
Span Setting Range:
Zero-Span and 1 MHz – 1000 MHz (1-2-5 Sequence)
Marker:
Frequency Resolution:
1 kHz, 6½ digit,
Amplitude Resolution:
0.5 dB, 3½ digit
Resolution Bandwidths
(RBW) @ 3dB:
500 kHz and 20 kHz
Video fi lter (VBW):
4 kHz
Sweep Time:
20 ms
Amplitude Characteristics (Marker Related) 150 kHz – 1 GHz
Measurement Range:
–100 dBm to +10 dBm
Scaling:
10 dB/div.
Display Range:
80 dB (10dB/div.)
Amplitude Frequency Response (at 10 dB Attn., Zero Span and RBW
500 kHz, Signal – 20 dBm):
±3 dB
Display (CRT):
8 x 10 division
Amplitude Scale:
logarithmic
Display Units:
dBm
Parameter Display (LCD):
2 lines, 20 characters each, Center Frequency, Span
Marker Frequency, Reference Level, Marker Level
Input Attenuator Range:
0 – 40 dB (10 dB increments)
Input Attenuator Accuracy
rel. to 10 dB:
±1 dB
Max. Input Level (continuous)
10 - 40 dB attenuation:
+20 dBm (0.1 W)
0 dB attenuation:
+10 dBm
Max. DC Voltage:
±25 V
Range of Reference Level:
–100 dBm to +10 dBm
Reference Level Accuracy rel. to 500 MHz, 10 dB Attn., Zero Span and
RBW 500 kHz:
±2 dB
Min. Average Noise Level:
approx. –100 dBm (RBW 20 kHz)
Intermodulation Ratio typical ›75 dBc (2 Signals: 200 MHz ,
(3rd Order):
203 MHz, – 3 dB below Reference Level)
Harmonic Distortion Ratio
(2nd harm.):
typical › 75 dBc (200 MHz, Reference Level)
Bandwidth Dependent Amplitude Error rel. to RBW 500 kHz
and Zero Span:
±1 dB
w w w . h a m e g . c o m
HM5510E/310505/ce · Subject to alterations · © HAMEG Instruments GmbH · ® Registered Trademark · DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 800 0 · Fax +49 (0) 6182 800 100 · www.hameg.com · info@hameg.com
Inputs / Outputs
Measurement Input:
Input Impedance:
VSWR (Attn.
10 dB):
Supply Voltage for Probes (HZ530): 6 V
Audio output (phone):
Test Signal output:
Frequency:
Level:
Functions
Keyboard Input:
Rotary Encoder Input:
Miscellaneous
CRT:
Acceleration Voltage:
Trace Rotation:
Ambient Temperature:
Power Supply:
Power Consumption:
Safety class:
Dimensions (W x H x D):
Color:
Weight:
Accessories supplied:
Spectrum Analyzer HM5510, Line Cord, Operators Manual,
2 x Adapter N to BNC
Optional accessories: HZ520 Antenna; HZ530 Near fi eld
Probe Set for EMI Diagnosis; HZ575 Converter (75Ω
50Ω)
A Rohde & Schwarz Company
S p e c i f i c a t i o n s
N-socket
50 Ω
typ. 1.5 : 1
DC
3.5 mm Ø jack
N-socket, output Impedance 50 Ω
10 MHz
0 dBm (±3 dB)
Center Frequency, Reference
Center Frequency, Reference Level, Marker,
Intensity (CRT), Contrast (LCD),
D14-363GY, 8 x 10 cm with internal graticule
approx. 2 kV
adjustable on front panel
+10° C to +40°C
105 - 253 V, 50/60 Hz ± 10%, CAT II
approx. 31 W at 230 V/50Hz
Safety class I (EN61010-1)
285 x 125 x 380 mm,
with adjustable, lockable tilt handle
techno-brown
approx. 5.6 kg
Subject to change without notice
25

Advertisement

Table of Contents
loading

Table of Contents