Safety Measures At The Installation Location - BARTEC BENKE DPA-4.5 Operating Manual

Distillation process analyzer
Table of Contents

Advertisement

Distillation Process Analyzer DPA-4.5
Safety
Eye protection
Sticker 9 (see p. 35)
Hearing protection

2.6 Safety measures at the installation location

Work room requirements
Containers and hoses
Vent/drain system
38
This sign indicates that suitable safety goggles must be worn
(see also p. 43).
This sign indicates that suitable ear protectors must be worn (see
also p. 43) and is located on the control box (when the control box
cooling option is installed).
Ensure that the work room is sufficiently ventilated.
Adhere to emission limit values. If necessary, install an exhaust
air treatment system.
Do not direct drawn-off air back into the work room.
Install washing facilities and eye baths in the workplace.
Install solvent-resistant flooring.
Creeping vapors can constitute an ignition source. Seal the
room properly to ensure that gases or vapors which could
contain ignition sources cannot enter the room.
Install equipment for detecting and reporting gas hazards.
Containers and hoses should be clearly labeled.
Only conduct work on containers and hoses when they have
been thoroughly rinsed and rendered inert.
Protect them against impermissible external heat sources.
Check for leaks regularly.
Do not feed any external gases, vapors or liquids into the
vent/drain system.
December 2013