Table 5. Wire Gauge and Distance to Remote
Alarm
Wire
Total Wire
Gauge
Length (feet)
20
530
18
840
16
1330
14
2120
12
3370
To install the remote alarm, make the following connections:
1. Connect the COMMON terminal on the cabinet switch to
the COMMON wire on the alarm.
2. To get an alarm when the switch contacts open, connect
the OPEN ON FAIL terminal on the cabinet to the OPEN
ON FAIL wire on the alarm.
3. To get an alarm when the switch contacts close, connect
the CLOSE ON FAIL terminal on the cabinet to the
CLOSE ON FAIL wire on the alarm. The COMMON and
CLOSE ON FAIL wires must be tied together in this
application
4. Plug the alarm system service cord into an electrical
outlet.
5. The contacts will trip in the event of high temperature
alarm or low temperature alarm.

Final Checks

Before start up, be sure to complete the following steps:
1. Make sure that the unit is free of all wood or cardboard
shipping materials, both inside and outside.
2. Check the positions of the shelves to adjust the positions.
3. Verify that the unit is connected to a dedicated circuit.
General Purpose Freezers
Distance to Alarm 1/2
Wire Length (feet)
265
420
665
1060
1685