eXact Chlorine Plus Instruction Manual page 8

Ideal for drinking water, pools, spas, environmental, food & beverage water quality testing

Advertisement

Peracetic Acid Test Procedure
PA5
REMOVE STRIP
1
Remove one (1) eXact
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: PA5
3
Press and re-press the MENU button until the display shows the parameter PA5.
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
the water sample.
ZERO METER
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
Sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
®
Dip the eXact
READ. This starts the 20 SECOND countdown timer. During this time move the strip in a gentle
back and forth motion (approx. 2 strokes/sec). Remove and discard the strip after "1" on the
display disappears. The cursor will move across the display while the meter prepares to measure
the sample. Record result displayed (this result is automatically stored in PA5). After testing, rinse
cell immediately.
Hydrogen Peroxide Test Procedure
HP6
REMOVE STRIP
1
Remove one (1) eXact
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
SELECT TEST: HP6
3
Press and re-press the MENU button until the display shows the parameter HP6.
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
the water sample.
ZERO METER
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
DIP STRIP -
6
®
Dip the eXact
READ. This starts the 20 SECOND countdown timer. During this time move the strip in a gentle
back and forth motion (approx. 2 strokes/sec). Remove and discard the strip after "1" on the
display disappears. The meter will begin counting up for 100 seconds, at the end of which, the
cursor will move across the display while the meter prepares to measure the sample. Record
result displayed (this result is automatically stored in HP6). After testing is completed, rinse cell
immediately.
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.
®
Strip Micro PA, Part No. 486675 from the bottle before
Strip Micro PA, Part No. 486675 into the CELL and immediately press
®
Strip Micro HP, Part No. 486648 from the bottle before
(read carefully and follow procedure closely)
Strip Micro HP, Part No. 486648 into the CELL and immediately press
8
PA
HP