Use

Positioning the microscope

Adjusting the mid position:
Press the "Reset XY-unit" key (4).
¸
¸
The XY unit will move to its middle position.
LED (2) is lit when the mid position has been reached.
1
9
Coarse positioning
Hold both handles (6) to position the microscope.
¸
¸
Fine positioning
Actuate the XY unit via the footswitch (7).
¸
¸
Adjust the travel speed of the XY unit on the control panel
¸
¸
(see page 21).
Tilt fine adjustment:
Use the tilt head fine adjustment (1) to set the tilt of the
¸
¸
microscope precisely.
18
2
4
6
+
7
8
Leica M620 F20 / Ref. 10 714 371 / Version 00
Auto reset:
The "auto reset" function can be deactivated on the
control panel in the service area.
Hold both handles and move the microscope to its
¸
¸
uppermost position.
All drives will move to their reset positions and the illumination
will go out. When the microscope is returned to its working
position, the illumination will switch on at the configured
starting value.
3

Adjusting the focus

(in accordance with factory settings)
Basic setting:
Press "reset focus" button (5).
¸
¸
The focus will travel to its home position (1/3 up, 2/3 down)
5
at maximum speed.
LED (3) is lit when the focus reset position has been reached.
Adjusting the focus:
Press the programmed button (8) or (9) on the footswitch.
¸
¸
Adjust the speed on the control panel (see page 21).
¸
¸

Adjusting the magnification

Adjust the magnification on the control panel (see page 20).
¸
¸