Internal Preamplifier Installation Kit
Introduction
This kit contains all of the instructions required to install the internal preamplifier options into an
X-Series signal analyzer. When Options P26, P44, P4L, P50, or P5L are installed, both the
Preamplifier and a Low Noise Amplifier (LNA) are licensed. Operation of the LNA with Option P5L is
limited above 43.5 GHz.
4
1. At the time of manufacture the hardware related to this option was fully
adjusted and the option performance was verified to be within its warranted
specifications. Within one year of the initial calibration date of the instrument
this option is fully calibrated with no further adjustment or verification testing.
To determine the initial calibration date, locate the original calibration
certificate that was shipped with the instrument at the time of purchase. The
Date of Calibration is printed on the original calibration certificate.
To ensure that this newly installed option has been installed properly, the
procedure that follows includes a functional check.
No factory test data is available for this option. If you require test data for this
option then a full calibration is required after installation of this option.
Arrangements regarding the level of calibration must be made with the
calibration provider.
2. The latest revision of the X-Series signal analyzer software may be
downloaded from:
http://www.keysight.com/find/Xseries_software
3. This option is licensed for one instrument model/serial number combination.
The license key will only install on the designated instrument.
Installation Note N9042-90005