Directives, Standards And Guidelines; Refrigerants; Table A-2: Directives, Standards And Guidelines Cryofuge 16 Heavy Duty Water Cooled; Table A-3: Refrigerants Used In The Cryofuge 16 Heavy Duty Water Cooled - Thermo Scientific Cryofuge 16 Instruction Manual

Heavy duty water cooled centrifuge
Hide thumbs Also See for Cryofuge 16:
Table of Contents

Advertisement

Thermo Scientific Cryofuge 16 Heavy Duty Water Cooled Centrifuges

1. Directives, Standards and Guidelines

Centrifuge
Region
Europe
Thermo Scientific Cryofuge
16 Heavy Duty Water Cooled
Single Phase
220 / 230 V, 50 / 60 Hz
USA & Canada
Single Phase
208 / 230 / 240 V, 60 Hz
Japan
Single Phase
200 V, 50 / 60 Hz
Table A–2: Directives, Standards and Guidelines Cryofuge 16 Heavy Duty Water Cooled

2. Refrigerants

Centrifuge
Refrigerant
Thermo Scientific Cryofuge
R-449A
16 Heavy Duty Water Cooled
Contains fluorinated greenhouse gases in a hermetically sealed system.
Table A–3: Refrigerants used in the Cryofuge 16 Heavy Duty Water Cooled
Technical Specifications
Directive
93/42/EEC Medical Devices
2011/65/EU RoHS
Directive on the Restriction
of the use of certain
Hazardous Substances in
electrical and electronic
equipment
Protective goals of:
2006/42/EC Machinery
2014/35/EU Low Voltage
2014/30/EU
Electromagnetic
Compatibility (EMC)
FDA
Product Code KSO
Centrifuge, Blood-Bank For
In Vitro Diagnostic Use
Device Class 1
Quantity
Pressure
1.79 kg
34 bar
A–2
Standard
EN 61010-1
IEC 61010-2-020
EN 61010-2-101
EN 61326-1 Class B
EN 61326-2-6
EN 62304
EN 62366
EN ISO 14971
EN ISO 13485
ANSI/UL 61010-1
IEC 61010-2-020
IEC 61010-2-101
IEC 61326-1 Class B
IEC 61326-2-6
EN 62304
EN 62366
EN ISO 14971
EN ISO 13485
IEC 61010-1 3
IEC 61010-2-020 3
IEC 61010-2-101 3
IEC 61326-1 Class B
IEC 61326-2-6
EN 62304
EN 62366
EN ISO 14971
EN ISO 13485
GWP
CO
e
2
1397
2.50 t

Advertisement

Table of Contents
loading

Table of Contents