Troubleshooting; Mechanical Emergency Door Release; Figure V-1: Marking On Windshield Lid - Thermo Scientific Cryofuge 16 Instruction Manual

Heavy duty water cooled centrifuge
Hide thumbs Also See for Cryofuge 16:
Table of Contents

Advertisement

Thermo Scientific Cryofuge 16 Heavy Duty Water Cooled Centrifuges

Troubleshooting

1. Mechanical Emergency Door Release

Serious injuries can occur, if you touch a spinning rotor with your hands or tools.
WA R N I N G
A rotor can still be spinning after a power failure occurs.
Do not open the centrifuge before the rotor has stopped spinning. Do not touch a spinning rotor.
Never use your hands or tools to stop a spinning rotor.
During a power failure, you will not be able to open the centrifuge door with the regular electric centrifuge door release. A
mechanical override is provided to allow sample recovery in the case of an emergency. The mechanical override should be
used only in emergencies and after the rotor has come to a complete stop.
Always wait until the rotor has come to a stop without braking. The brake does not work without power supply. The
deceleration process lasts much longer than usual.
Proceed as follows:
Wait until the rotor has stopped. It can take longer than 40 minutes.
1.
Use the sight glass in the centrifuge door to make sure that the rotor has stopped. The handle of the windshield lid has
a marking, that shows a semicircle, if the rotor has stopped.
Figure V–1: Marking on windshield lid
Troubleshooting
V
V-1

Advertisement

Table of Contents
loading

Table of Contents