14 Error messages

ERR01
ERR02
ERR03
ERR04
ERR05
ERR06
ERR07
ERR08
"UNLOAD"
"CAL But"

15 Maintenance, service and disposal

Before you start any works related to the maintenance, cleaning and
repair, disconnect the device from the operating voltage.

15.1 Cleaning

Do not use any aggressive cleaning agents (solvents etc.), but clean the device with
a cloth and mild soap solution. The liquid must not get inside the device. After the
cleaning is finished, wipe the device dry using a soft cloth.
Any loose specimen/powder remains can be removed carefully with a brush or a
handheld vacuum cleaner.
Remove any scattered weighed material immediately.

15.2 Maintenance and service

 The device can be operated and maintained solely by the technicians trained
and authorized by KERN.
 Disconnect from the mains before opening.
TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2011
Unstable weight value or impossible resetting.
Check environmental conditions.
Adjustment error, e.g. unstable environmental conditions.
Adjustment error, e.g. incorrect adjustment weight.
Too low/unstable unit weight.
Data transmission impossible as the weight value is unstable.
Check environmental conditions.
Unstable weight value in the density determination mode.
Check environmental conditions.
Erroneous data upload (counting the number of pieces, density
determination etc.).
Internal adjustment error.
Weighing range exceeded.
Check the scale plate position.
Carry out scale adjustment.
Weighing range exceeded (upwards), the load placed exceeds
the scale weighing range. Remove the load from the scale.
Weighing range exceeded (downwards), e.g. the scale plate
not placed.
95