Test Weighing - KERN ALJ Operating Manual

Analytical and precision balances
Hide thumbs Also See for ALJ:
Table of Contents

Advertisement

12.4 Test weighing

The <Test weighing> application enables to determine the upper and lower limit
value and, consequently, to ensure the weight of the weighed material belongs to the
range between the determined tolerance limits.
The tolerance symbol () and an audible signal (possible choice) indicate if the
weighed material belongs to the range between two tolerance limits.
Result presentation
1. Lower and upper limit value ≥ 0
Tolerance
symbol
2. Lower limit value > 0, upper limit value = 0
Tolerance
symbol
3. Lower limit value = 0, upper limit value > 0
Tolerance
symbol
TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2011
Audible
Assessment
signal
no
Weighed material above the preset tolerance
yes
Weighed material in the preset tolerance range
no
Weighed material below the preset tolerance
Audible
Assessment
signal
yes
Weighed material > the lower limit value
Weighed material ≤ the lower limit value
no
Audible
Assessment
signal
no
Weighed material > the upper limit value
Weighed material ≤ the upper limit value
yes
81

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents