Buzzer Set Up - Labnet I-4001-HCS Instruction Manual

Microtube shaking incubator
Hide thumbs Also See for I-4001-HCS:
Table of Contents

Advertisement

TEMP
TIME
SPEED
TEMP
TIME
TEMP
TIME
TEMP
TIME
TEMP
TIME
Press both
Speed and

Speed at the same time,
and the display will show
BEEP.
TEMP
TIME
Press both
Temp and

Temp at the same time
to start the calibration. The
TEMP
display will show 40.0 and
the temperature of the
TIME
instrument will automatically
increase to 40°C. Once the
temperature is kept constant,
timing will start.
TEMP
TIME
TEMP

8.5 Buzzer set up

TIME
Upon failure's occurrence or ending, the instrument will buzz. You may disable it if you do not need

this function. However, the default setting is to have it on.
This instrument has keyboard beep. Once the button is pressed, it will beep once. You may disable it if

you do not need this function. However, the default setting is to have it on.
Press Short/Prog for once

and change ON to OFF.
TEMP
TIME
30 minutes later, press

Short/Prog for one time,
and the display will show the
blink of TEMP. At this time,
the display will show the
actual temperature reading
of the thermometer. For
example, if the thermometer
reads 40.5°C, you can adjust
the temperature display
to 40.5°C by pressing the
temperature set up button.
Next press Short/Prog for
one time, and the display
will show 100.0°C. Following
the same instruction, when
the temperature of the
TEMP
instrument increases to 100°C,
wait for 30 minutes, input
TIME
the calibrated value and press
Short/Prog to save it.
TEMP
TIME
Press
Speed and

Speed at the same time
to save the setting, and
buzz will be disabled upon
failure's occurrence or
ending. The display will
then show the keyboard
beep status. If the display
will show ON K_S. Press
Short/Prog once and
change ON to OFF.
TEMP
TIME
When the calibration

of 40°C is done, press
Start/Stop to exit. At this
time the calibration of
TEMP
40°C is ineffective. When
the calibration is done
TIME
TEMP
for once at both 40°C
and 100°C , check for
TIME
the discrepancy between
the two calibration
points and the actual
temperature points.
It should be within
0.5°C. If it's more than
TEMP
0.5°C, please follow the
above instruction and
TIME
recalibrate until the
request is met.
Then press both

and
Speed at the same
time to save the setting,
and keyboard beep will
be disabled. If you want
to exit during set up, you
may press Start/Stop
and the setting will be
ineffective.
Instruction Manual
TEMP
TIME
Speed
|
7