Labnet VorTemp 56 Instruction Manual

Labnet VorTemp 56 Instruction Manual

Shaking incubator
Hide thumbs Also See for VorTemp 56:

Advertisement

Quick Links

Labnet VorTemp™ 56 Shaking Incubator
Instruction Manual
Catalog Numbers:
S2056A
S2056A-230V

Advertisement

Table of Contents
loading

Summary of Contents for Labnet VorTemp 56

  • Page 1 Labnet VorTemp™ 56 Shaking Incubator Instruction Manual Catalog Numbers: S2056A S2056A-230V...
  • Page 2: Table Of Contents

    Table of Contents 1.0 Introduction ..................1 1.1 Symbols and Conventions .
  • Page 3: Introduction

    1.0 Introduction The Labnet VorTemp™ 56 shaking incubator is a benchtop shaker and heating chamber for incubating and shaking sensitive samples. It replaces two devices, reducing both time and space needed. It is suitable for biochemistry, microbiology, and clinical laboratories in which applications require temperature and shaking treatment.
  • Page 4: Construction

    3.1. Construction The housing of Labnet VorTemp™ 56 shaking incubator is constructed of steel plate coated with highly resistant polyurethane lacquer. The interior chamber is isolated with special thermal protection materials and insulation foam. Both the shaking mechanism and temperature chamber regulated via microprocessors, which control all sensors for motor speed, temperature, and time.
  • Page 5: Microplate Holder Platform

    Insert the microplates with your samples, and start the device. Screw Unscrew NOTE: When you attach the microplate platform, it must be attached so that the arrows are pointed towards the front of VorTemp 56 shaking incubator. Instruction Manual...
  • Page 6: Safety Information

    Switch on the device using of the ON/OFF switch on the front panel. The switch will illuminate to indicate that the power is On.  Check the running parameters, and set new parameters if necessary. Refer to the instructions described in Section 7.  LABNET VORTEMP 56 SHAKING INCUBATOR...
  • Page 7: Operating Instructions

    7.0 Operating Instructions 7.1. Control Panel Encoder Time [min] Power 88.8 RPM [x10] Start Stop Temp [°C] 1. Power Key: Switch ON (illuminates when on) or OFF. 2. Run: Green signal light illuminated when the shaker is operating. 3. Time: Yellow signal light illuminated when the unit is set to adjust time. 4.
  • Page 8 Rotating the Encoder knob right (+) or left (-) sets the temperature to the desired value:  37.4 = 37.4°C If you want to use the incubator without temperature control activated, turn the Encoder knob under 0.5 or above 99.5 until the  display reads OFF. LABNET VORTEMP 56 SHAKING INCUBATOR...
  • Page 9: Shaking Without Heating

    Pressing the Start/Stop Button Run and Time are illuminated. The timer will count down from the set time value.  NOTE: You cannot adjust the set time during shaking. Encoder Time [min] Power 99.4 RPM [x10] Start Stop Temp [°C] RMP Adjusting During a Run Push the Encoder knob until the RPM light is illuminated.
  • Page 10: Additional Operations

    8.0 Troubleshooting The Labnet VorTemp™ 56 shaking incubator features built in self-diagnostic procedures which are constantly checking the operating parameters and performance, as well as functions that are necessary for safe and reliable operation. An error code is shown on the LCD display if an error occurs.
  • Page 11: Maintenance And Cleaning

    calibrated thermometer with precision 0.1°C or more. After allowing the temperature to equilibrate for 2 hours, read the temperature on the thermometer and compare it with the temperature on the LED display. This difference between the thermometer and the display is the value which you will enter into the unit to recalibrate it. Example 1: Temperature on the thermometer is 37.9°C, temperature on display is 37°C.
  • Page 12: Equipment Disposal

    12.0 Equipment Disposal According to Directive 2012/19/EU of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment (WEEE), this product is marked with the crossed-out wheeled bin and must not be disposed of with domestic waste.

This manual is also suitable for:

S2056aS2056a-230v

Table of Contents