5. Analysis of Test Sticks

5.1 How to perform a measurement

The instrument is very easy to use. In order to start the measurement, the
test stick is placed on the strip holder. The instrument automatically detects
a new strip and starts the measurement. A progress bar appears, that
indicates the remaining analysis time. After 30 seconds the test stick is
drawn into the instrument, after 60 seconds it is released.
Notice:
Make sure to remove excess urine by blotting the test stick carefully
on a lint-free cloth.
Notice:
If auto mode (Chapter 9.5 Autostart) is deactivated, the analysis
needs to be started by pressing
After the measurement, the instrument will release the analysed test stick
wich can now be discarded. The result is displayed on the screen and is
transferred via the interfaces and / or printed according to equipment
settings.
For additional information on the test strip, please read the package insert
that comes with the strips.
DANGER OF INFECTION:
Urine and used test sticks bare the danger of infection. Always use
protective gloves during handling and disposal. The disposal of used
test sticks should be performed according to the regulations for the
handling of potentially infectious material.
Manual EN V2.00 / 06-2009
on the touch-screen.
18