Schiller AT-102 Service Handbook page 49

12-channel ecg recorder
Table of Contents

Advertisement

AT-102 ECG Service Handbook
System Settings
Filter
Interpretation
Comm
Spiro
Stress
Date and Time
Article Number 2. 540 028
SCHILLER AG 2002
Baseline Filter
0.05, 0.15 or 0.30 Hz
Mains Filter
50, 60 Hz or OFF (-)
Myogram Filter
25 or 35 Hz, ON (+) or OFF (-)
SSB Filter
Smoothing Filter Enabled (+) or Suppressed
(-)
SSF Filter
Baseline Stabiliser Enabled (+) or
Suppressed (-)
Sensitivity
Low (-) or high (+) sensitivity
A30
Patient age is assumed to be < 30 (-)
or >30 (+)
U
'Unconfirmed report' is written (+) or
suppressed (-)
Abnormal
Normal / Abnormal printed Enabled (+) or
Suppressed (-)
Baud rate
115200, 57600, 38400, 28800, 14400 or 9600.
This is followed by parity setting (Y/N), bits
and number of stop bits.
Mode
Line or Modem
F=f(v)
Enabled (+) or Suppressed (-)
Flow =f(V)
Enabled (+) or Suppressed (-)
Diagn.
Enabled (+) or Suppressed (-)
PEFD (l/min)
Enabled (+) or Suppressed (-)
Axis
5, 10 or 20 mm/s
Bike
900 / 911 or SECA CT100 or none
Prot
252525 or user
TM
ETM435/TM4000ES or RAM 770CE or MTM-
1500 or none
Prot
Bruce or user
BP
Manual, automatic of none
Stage Print
leads and format
ST Lead
Defined lead for ST Measurement
J- Point
J plus 20ms, 40ms, 60ms, or 80ms
System date and time, and user ID (if set)
Section 1
Operating Elements
1.31

Hide quick links: