AT-102 ECG Service Handbook
Upgrading the Unit / Updating the
Software
5.
Install the SWUP program on the PC (available from the SCHILLER Mailbox).
6.
Follow the instructions given by the program.
7.
Open the SWUP program through the SWUP icon on the desktop or via the Startup -
Schiller-Menu. The following is displayed:
8.
Click the config icon. The following screen is displayed:
9.
Select the applicable COM port (1 or 2). Set the Baudrate to `Auto` (or to the same
Baud rate as set in the AT-102.
10.
Click on `update version` to update the system software.
11.
A series of beeps indicates that the software has been successfully updated. The
updated software can be used immediately.
Article Number 2. 540 028
SCHILLER AG 2002
Section 5
Adjustments
To exit the
program
without
updating
click the
close icon
5.17