Configuring The Programmable Output - ABB VA Master FAM540 Commissioning Instructions

Metal cone variable area flowmeter
Hide thumbs Also See for VA Master FAM540:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 29
5.5

Configuring the programmable output

The contact output of the transmitter is configured by default as a NAMUR contact:
The contact can be changed, if necessary.
Fig. 20
1
A
B
Important
For explosion-proof devices, remove the safety locking device when opening the cover and
reattach it after closing the housing.
CI/FAM540-X1
Contact for switching the output configuration
Switch on left stop: NAMUR contact
Switch on right stop: Optocoupler function
VA Master FAM540
Startup Operation
EN - 23