5.6

Calibration Menu

The calibration screen is where users calibrate sensors, view calibration records, and set calibration reminders. This section will
explain the calibration options and settings. Information related to calibration methods for a specific parameter calibration can be
found in
Section
4.
Calibrate
This displays a list of parameters that available to calibrate. The parameters are organized under each respective sensor. Every
parameter has two options:
1. CALIBRATE – Select this to perform a user calibration.
2. FACTORY RESET CALIBRATION – Select this to restore the factory default calibration. Note this deletes the user calibrations from
the sensor and reverts to the original factory settings. A user calibration must be performed after the factory reset.
Find Calibration Records
This opens the calibration records database where users can filter and find previous calibration records. A calibration record is
generated and stored every time a parameter is calibrated. Multiple calibration records may be selected to view simultaneously.
Selected records are listed under the Calibration Records Panel. These records are sorted by calibration date and organized by
sensor on the left side of the screen. Select a specific record to view its calibration details displayed on the right side of the screen.