Sdi-12 Setup; Output Configuration - Xylem ProSwap 610224-01 User Manual

Table of Contents

Advertisement

7.1

SDI-12 Setup

The ProSwap Logger includes native SDI-12 output for use with flying lead cables for a direct interface into a 3rd party data collec-
tion platform (DCP). Refer to the wiring diagram below for connecting the cable to a terminal:
Ground
SDI-12 on
Datalogger
Regulated 12VDC
power supply
(not included)
7.2
SDI-12 Output Configuration
In order to appropriately setup a sonde to communicate measurements to a datalogger, it is critical to align the settings from the
sonde and the logger.
In Kor software, go to the Deployment menu and either open a template or create a template from the sonde. Go to the DCP
Adapter Output tab, and select the parameters and order of output.
The complete list of parameters are listed in the "Available" box. Move the desired parameters to the "Selected" parameters box
and organize them using the arrow buttons. See
Once the parameter output set, save and apply the template to the sonde. This template can be saved locally on the PC, but it must
also be pushed to the sonde for the settings to take effect. So be sure to apply the template to the sonde.
NOTE:
There are two options when applying the template to the sonde, apply without logging or with logging. Either option
may be used. When deploying with logging the sonde will create a redundant log file inside the sonde. To ensure
internally logged data matches output data, be sure to select "Sample and Hold" for sonde's logging mode in the
Advanced tab of the deployment template. Without logging the data will only be available through the SDI-12
output.
Signal Output
Black Wire
Bare Wire
Orange Wire
Section 5.8 DCP Adapter Output
Flying Lead Adapter
Cable to ProSwap Logger
Red Wire
for more information.