FLIR A6000 Series User Manual page 75

Table of Contents

Advertisement

The provided breakout cable has numerical markings on the BNC overmolding that are described in
the table below.
Pin
Name
1
NC
2
NC
3
RECORD
START
4
SYNC OUT
5
GROUND
6
GPIO IN
7
Reserved
8
TRIGGER
LOCK-IN
(OPTIONAL)
[A6260/A6750
only]
9
INTEGRATE
ACTIVE
10
SYNC IN
11
RS-232 TX
12
RS-232 RX
Inputs are all LV-CMOS. High>2V, Low<0.2V. Max is 5.5V
If you wish to make your own breakout cable, there are three variants of the Hirose mating connector
that will work: HR10A-10P-12S(73), HR10-10P-12S(73) or HR10A-10P-12SC(73).
A6000 and A8500 Series User's Manual
Breakout
Type
Connector
BNC 3
INPUT
BNC 4
OUTPUT
BNC 6
INPUT
BNC 7
BNC 8
INPUT
BNC 9
OUTPUT
BNC 10
INPUT
DB9 Male
OUTPUT
DB9 Male
INPUT
Description
Sets RecordStart bit in image header. Can be used
by FLIR Research Studio to start recording.
20 microsecond wide pulse at start of integration
8-bits. Sets bits in image header. Reflected in the
"DigitalIn header field (hex). Updated at ~1Hz rate.
External trigger. Can be used by FLIR Research
Studio to start recording.
If lock-in option is enabled, replaces TRIG IN.
This output is either high or low (user selectable)
when the camera integration is occurring.
The Sync In can be selected, by the user, to operate
as an external clock. It is a rising edge LV-CMOS
signal (5.5V max). The minimum width is 160nS.
For camera control
For camera control
6 – Interfaces