  • Page 1 English OPERATION MANUAL Version ecom-UNO 200 hPa Status Nov. 2020...
  • Page 2: Intended Use

    Intended Use The ecom-UNO is suitable for measuring the positive, negative and differen- tial pressure of gases. It is particularly suitable for work on heating, venti- lation and air conditioning systems. The measuring range covers pressures from -200 hPa to +200 hPa.
  • Page 3: General Safety Instructions

    General Safety Instructions WARNING Magnetic Field The magnetic field can cause pacemakers to malfunction in their impair. Operation of the ecom-UNO by persons with pacemakers is prohibited. WARNING Hazard of Injury Improper work on incinerators can cause injuries. Work on incinerators may only be done by qualified specialists.
  • Page 4: Device Description

    Device Description Connection Overpressure Battery Capacity HOLD Key Connection Depression Cursor Setting Key Display Pressure Units ON/OFF...
  • Page 5: Magnetic Holder

    Magnetic Holder The ecom-UNO has a magnetic holder on the back, with which the device can be attached to most metal surfaces. WARNING Magnetic Field The magnetic field can cause pacemakers to malfunction in their impair. Operation of the ecom-UNO by persons with pacemakers is prohibited.
  • Page 6 • The zero point correction is used to compensate the zero point drift after switching on the instrument (see setup). Range Switch • If the pressure of 0.5 hPa/mbar is exceeded, the ecom-UNO auto- matically switches to the higher measuring range. • Resolution from 0.001hPa/mbar to 0.01 hPa/mbar and from 0.01 hPa/mbar to 0.1 hPa/mbar.
  • Page 7 • Maximum value, the highest value since the device was switched on, is displayed. 4. Press the HOLD key 4 times. • ecom-UNO is in measuring mode again. 5. Hold the HOLD key. • Minimum and maximum values are deleted.
  • Page 8: Change Batteries

    Change Batteries NOTICE The two 1.5 V AA batteries must be replaced when the battery capacity indicator flashes. 1. Open the lid of the battery compartment. 2. Remove used batteries. 3. Replace batteries with new batteries. 4. Slide the cover onto the battery compartment. Setup Settings can be made in the setup menu.
  • Page 9: Technical Data

    Technical Data Measurement Range 200 hPa Resolution 0.01 hPa Overload 300 hPa Accuracy ±1 % of the MBE Ambient Temperature 0 °C – 40 °C Storage Temperature -20 °C – 70 °C Batteries 2x 1.5 V AA Power Supply 20 mA Size 106 mm x 64 mm x 28 mm Weight...
  • Page 10: Maintenance

    Maintenance To ensure the accuracy of your measuring device, we recommend an annu- al inspection by an authorised ecom partner. In case of heavy use, shorter inspection intervals should be chosen – please contact your ecom partner. Disposal • Dispose of used batteries at the designated collection points.
  • Page 11 Consumables/Spare Parts NOTICE You can find consumables or spare parts in our webshop at www.shop.ecom.de.
  • Page 12 GmbH Am Großen Teich 2 D-58640 Iserlohn info@ecom.de www.ecom.de Phone: +49 2371 945-5 Fax: +49 2371 40305...