Check the cabling by operating the motors of the bed to their outer positions. When all wires have been correctly
installed, there is no squeezing risk of the cables.
It is normal for cables to loosen slightly after short usage.
The cables are equipped with plugs at both sections, hence, they can be replaced independently
(except for the height motor).
The cables are attached to the top frame with cable clips.
The hand control, control unit and motors are protected according to IPX4 or IPX6.
It can be dangerous to roll over the mains cable.
Do not bring mains cable into moving parts.
Disconnect the plug from the mains before moving the bed.
The cables must be mounted in such a way that they are kept clear of the floor and do not
block the castors.
A lock cam must be used on the control unit - if not, Invacare®cannot guarantee the IP
protection.