7.4 Maintenance chart
Service and maintenance of the Invacare® Octave
received the necessary instruction or training.
S/N (located on the mattress support): ___________________
Date:
Circlips and cotter pins checked.
Screws tightened.
Weldings checked.
Side rail locking and moving system tested.
Castor fittings tightened.
Castor brakes checked.
Height adjustment motor checked.
Back rest motor checked.
Thigh section motor checked.
Leg section motor checked.
Wiring correct and undamaged. Plugs undamaged.
Damaged paintwork repaired.
Lubrication according to the lubrication plan
(chapter 12).
The wooden side rails´ gliding system must not be
lubricated with oil - otherwise the wooden bars will
move sluggishly.
Accessories checked.
Measurements can be done at internal threads after removal of finger screws.
7.5 Motors, control unit and hand control
These parts are not serviceable, they must be replaced.
A safety test comprising control of the motors' performance and mechanical state is recom
mended once a year.
Motors, control unit and hand control must be regularly cleaned from dust and dirt and they must be inspected for
mechanical damage or breakage.
Inspect anchor points, cables, piston rod, casing and plugs and check the correct functioning of the motors.
TM
must only be performed by personnel who have
Initials.: