7. Cleaning and maintenance

The base frame must be supported during service inspections to prevent accidental lowering
Only personnel who have received the necessary instructions or training may perform service and mainte-
nance of the bed.
With normal daily use, service must be carried out according to the service schedule after 2 years use and
thereafter every second year.
For beds with electrical equipment a safety test comprising control of the motors' performance and
mechanical state is recommended once a year.
When reconditioning the bed, service must be carried out according to the service schedule.
7.1 Cleaning
The bed does not tolerate cleaning in automatic washing plants or the use of water-jet based cleaning equipment.
1.
Please make sure the power plug is removed from the wall socket outlet before cleaning.
2.
The bed should be washed down using a wet sponge, cloth or brush.
3.
Use ordinary household cleaning agents. Never use acids, alkalines or solvents such as acetone or cellulose
thinner.
4.
The backrest and the leg/thigh sections can be lifted from the mattress support, which facilitates cleaning and
access to the control unit.
5.
The hand control, motors and control unit may be washed with brush and water.
Important details for cleaning IPX6 and IPX4 hand controls, control unit and motors:
Water may be pressurised, but never use high-pressure cleaning or steam cleaning directly on the electrical
parts.
6.
Dry the bed after cleaning.
Make sure that all electric plugs are in place during washing
.