9.1 Guidance and manufacturer's declaration - electromagnetic emission
Guidance and manufacturer´s declaration – electromagnetic emission
The Octave Bed is intended for use in the electromagnetic environment specified below. The customer or
the user of the Octave Bed should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer´s declaration – electromagnetic immunity
The Octave Bed is intended for use in the electromagnetic environment specified below. The customer or the
user of the Octave Bed should assure that it is used in such an environment.
Immunity test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
± 2 kV for power
Electrostatic
supply lines
transient / burst
± 1 kV for input/output
IEC 61000-4-4
lines
± 1 kV differential
Surge
mode
IEC 61000-4-5
± 2 kV common mode
< 5 % U
(>95 % dip in U
for 0,5 cycle
Voltage dips,
40 % U
short
(60 % dip in U
interruptions and
for 5 cycles
voltage variations
on power supply
70 % U
input lines
(30 % dip in U
for 25 cycles
IEC 61000-4-11
< 5 % U
(>95 % dip in U
for 5 sec
Power frequency
(50/60 Hz)
magnetic field
3 A/m
IEC 61000-4-8
NOTE
U
is the a. c. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
The Octave Bed uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment. The
Octave Bed is suitable for use in all establishments, including
Class A
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Complies
IEC 60601
Compliance level
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 1 kV differential
mode
± 2 kV common mode
< 5 % U
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
< 5 % U
T
)
(>95 % dip in U
T
for 5 sec
3 A/m
Electromagnetic environment - guidance
Electromagnetic environment -
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
T
)
T
Mains power quality should be that
of a typical commercial or hospital
)
environment. If the user of the
T
Octave Bed requires continued
operation during power mains
interruptions, it is recommended that
T
)
the Octave Bed be powered from an
T
uninterruptible power supply or a
battery.
T
)
T
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.
22