  • Page 1 Invacare®Octave User's Manual (GB)
  • Page 2 Identification label and QA-mark are located on the bed confirming that the finished goods control has approved the bed. Please read the entire user’s manual before using the bed. Invacare® EC-Høng A/S is certified according to ISO 9001 and ISO 13485. Kind Regards René Engskov...
  • Page 3 Un’apposita targhetta e la marcatura QA sono state apposte sul letto a conferma dell’avvenuta verifica e accettazione del prodotto da parte del servizio qualità. Prima di utilizzare il letto vi invitiamo a leggere integralmente il manuale d’uso. Invacare® EC-Høng A/S è un’azienda certificata ai sensi della Normativa ISO 9001 e ISO 13485. CERTIFICACIÓN DEL PRODUCTO Enhorabuena por su nueva cama Invacare®Octave...
  • Page 4: Table Of Contents

    General Information ........8 Receiving Invacare®Octave ....... . 9 Disassembly parts .
  • Page 5: User Part

    A specially designed 4-sectioned mattress support provides the user with excellent comfort and provides optimal working conditions for the carer. In order to optimise the patient’s comfort Invacare® EC-Høng A/S recommends using a 15-cm mattress. (see order numbers sec. 8).
  • Page 6: Operation Of Invacare®Octave

    2. Operation of Invacare®Octave Sitting position (green button) Sitting position (green button) Down Button with the symbol shown Button with the symbol shown above. above. Push the left-hand side of the Push the right-hand side of the button. button. Backrest up...
  • Page 7 The castors might under special conditions be difficult to manouvre due to the floor covering being untreated or in poor condition. In matters of doubt, Invacare recommends the placement of a suitable type of protection between the castors and the floor.
  • Page 8 Adjusting the height of the lifting pole Loosen the cord as shown in fig. A. The lifting handle can now be adjusted to the desired height. Press the cord together as shown in fig. B and check that the cord is locked in the cord lock by pulling the handle.
  • Page 9: Technical Information

    If the functions of the bed change, please send the bed immediately for a check-up according to the maintenance chart, chapter 10. The Invacare®Octave must be stored in a room with a humidity equivalent to normal indoor conditions. The bed must be stored in a frost-free environment.
  • Page 10: Receiving Invacare®Octave

    5. Receiving Invacare®Octave Due to condensation, may the bed may not be used until it has gained room temperature. Check whether the bed shows any signs of damage. If the bed is damaged, see terms of delivery. The bed - mattress support and base frame - will be delivered assembled on a trolley.
  • Page 11 Connect Invacare®Octave to the mains. For ergonomic working position: Raise the bed (hi/low) to approx. 60 cm above floor level. Ref. chapter 2 “Operation of the Invacare®Octave”. 1. Assembling the fittings. a) Push bed end through the pipes on the top frame.
  • Page 12 2. Mounting the Britt IV wooden siderail (accessories). Raise the bed about one third. Remove the protective tape from the locking pins at the ends of the siderail. Mount one end at a time. Lift the lower wooden bar and guide the metal bow to the bed end rail (A). Press the locking pin of the siderail with a finger (B).
  • Page 13 4. Mounting the lifting pole (accessories). a) Push the lifting pole bottom piece in the fitting under the backrest. b) Mount the enclosed screws and nuts and tighten them. 5. Mounting the mattress support extension (10 cm). • It is not possible to use the wooden siderail in combination with mattress support extension. Therefore dismount the wooden side rails first.
  • Page 14: Disassembly Parts

    6. Disassembly parts Dismounting mattress support - base The mattress support can be disassembled from the base. To disassemble the mattress support from the base relevant ools and 2 persons. a) Dismount all accessories incl. bed end. b) Lower the bed into lowest position. c) Dismount the cable from the motor for height from the control unit.
  • Page 15: Wiring

    7. Wiring Note! Mains cable is placed below the frame of the base. The bed is not intended for use together with MEDICAL ELECTRICAL EQUIPMENT. It can be dangerous to roll over the mains cable. Do not bring mains cable into contact with moving parts. Disconnect the plug from the mains supply before moving the bed.
  • Page 16: Part Numbers For Accessories

    8. Part numbers for accessories Article Dimensions Order no. Britt IV wooden siderail (Beech, complete pair) ......... . 1427875-0101 Britt IV wooden siderail (Cherry, complete pair) .
  • Page 17: Service And Maintenance

    9. Service and maintenance Service and maintenance of the Invacare®Octave must only be performed by personnel who have received the necessary instruction or training. With normal daily use, service must be carried out according to the service schedule after 2 years use and thereafter every second year.
  • Page 18: Maintenance Chart

    10. Maintenance chart Service and maintenance of the Invacare®Octave must only be performed by personnel who have received the necessary instruction or training. Bed-serial/Id. no.: ___________________ Date: Initials.: Circlips and cotter pins checked. Screws tightened. Weldings checked. Siderail locking and moving system tested.
  • Page 19: Trouble-Shooting With The Electrical System

    Defective motor Motor sounds not move does not move 12. Lubrication plan We recommend to lubricate the Invacare®Octave according to the following instructions: Guideways - lubricate with grease Swivel on shear arms - lubricate with oil Swivel on mattress support -accessories...
  • Page 20: Cleaning

    13. Cleaning The Invacare®Octave cannot withstand cleaning in washing machines nor flushing with a water jet. Use ordinary household detergents. Dry the bed after cleaning. Never use acids, alkalines or solvents. Please make sure the power plug is removed from the socket.
  • Page 21: Disposal

    16. Disposal This product has been supplied from an environmentally aware manufacturer that complies with the Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/CE. This product may contain substances that could be harmful to the environment if disposed of in places (landfills) that are not appropriate according to legislation.
  • Page 22 Sweden & Finland Nederland Italy INVACARE AB INVACARE B.V. INVACARE MECC SAN S.R.L. Fagerstagatan 9 / Box 66 Celsiusstraat 46 Via dei Pini 62 S-163 91 Spånga NL-6716 BZ Ede I-36016 Thiene (VI) Phone: +46 8 761 70 90...